admin 管理员组文章数量: 1086019
2024年12月29日发(作者:decodeuri在线解码)
Product Insert
CatchPoint
™
Cyclic-GMP Fluorescent Assay Kit
Product # R8075
(Bulk Kit)
Quantity: 960, 120 µL reactions
Introduction
About the cGMP Assay
The CatchPoint
™
cyclic-GMP Fluorescent Assay Kit measures levels of 3’, 5’-cyclic
guanosine monophosphate (cGMP) or guanylate cyclase activity via a competitive
immunoassay for cGMP. The assay requires only a single washing step, and readings can
be taken in as little as 10 minutes or as long as 24 hours following substrate addition,
since no termination step is needed. The cGMP Fluorescent Assay Kit is suitable for use
in cell-based assays.
Principle of the Assay
The cGMP in the sample or standard competes with horseradish peroxidase (HRP)-
labeled cGMP conjugate for binding sites on the anti-cGMP antibodies (Fig. 1). In the
absence of cGMP, most of the HRP-cGMP conjugate is bound to the antibody. Increasing
concentrations of cGMP competitively decrease the amount of bound conjugate, thus
decreasing measured HRP activity.
HRP
HRP HRPHRP
HRPHRP
HRP
No cGMP Increasing cGMP
Maximum HRP activity
Decreasing HRP activity
cGMP
cGMP-HRP
HRP
Conjugate
Goat anti-Rabbit IgG
Rabbit anti-cGMP coated microplate
Antibody
Figure 1. Principle of the CatchPoint cGMP assay system
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
1
Applications
The kit is designed for use in applications where cGMP is generated in biochemical assay
systems, such as assays of guanylate cyclase activity. In addition, this kit can be used for
cell-based assays to measure intracellular cGMP levels following stimulation with peptide
hormones or stimulants such as nitric oxide (NO), sodium nitrate, and nitroprusside to
soluble or particulate guanylate cyclases.
Kit Components
Materials and Equipment
The following table lists the kit components.
Table 1: The CatchPoint cGMP Fluorescent Assay Kit (P/N R8075) contents
Reagent Quantity Description
96-well black plates 10 plates 96-well plate coated with Goat anti-
Rabbit IgG. Plates are pre-blocked and (clear bottom)
ready to use.
cGMP Assay Buffer 950 mL
Rabbit anti-cGMP Antibody 2 vials Antibody, lyophilized
cGMP Calibrator 2 vials
30 µM cGMP Calibrator, lyophilized.
On reconstitution with 5 mL assay
buffer each vial contains 30,000 pmol
cGMP / mL.
HRP-cGMP Conjugate 2 vials HRP-cGMP Conjugate, lyophilized
10X Wash Concentrate 950 mL
Cell Lysis Buffer 450 mL
Stoplight Red Substrate 2 x 900 µL
100X Substrate in DMSO
Substrate Buffer 950 mL
When the working concentrations of the above reagents are used as suggested, each kit
provides sufficient reagents for ten 96-well microplates with 120 µL total assay volume.
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
2
Storage and
All kit components are to be stored at 4°C.
Handling
IMPORTANT: Allow all the reagents to warm to room temperature prior to use.
Note: Small volumes of product will occasionally collect in the cap of the product vial during
shipment. Gently tap the vial on a hard surface or briefly centrifuge the vial to collect any liquid
trapped in the vial cap.
When stored properly, the kit components are stable for six months from the date of receipt.
The reconstituted working solutions of Rabbit anti-cGMP, cGMP Calibrator, and HRP-cGMP
Conjugate are stable for 3 weeks at 4°C. To insure optimal performance of the reagents do not
store below 4°C once reconstituted.
If the entire plate will not be used, protect any unused wells with a plate sealer and store at 4°C
in the original foil wrapper protected from light.
IMPORTANT: The reconstituted substrate solution is sensitive to light. To insure optimal
performance we recommend preparing a fresh stock solution (keep protected from light) and
adding to the assay plate within 60 minutes. Following substrate addition readings can be
taken at 10 minutes or as long as 24 hours.
Materials
Required but
not Provided
The following tables list the materials required but not supplied.
Table 2: Reagents and supplies
Reagent Item
Source
3% wt / vol hydrogen peroxide (H
2
O
2
) Major laboratory suppliers (MLS)
MLS
MLS
MLS
1.5 mL polypropylene capped
centrifuge tubes
10 mL capped centrifuge tubes
Pipettors – adjustable volume
Multi-channel troughs MLS
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
3
Equipment Item
Table 3: Compatible Instruments available from Molecular Devices (MDC)
Source
Analyst System
One of the following:
- Analyst GT
- Analyst HT
MDC P/N 0200-6003 or 0200-6004
MDC P/N 0200-6043 or 0200-6044
Gemini XPS
MDC P/N Gemini XPS Spectrophotometer
Gemini EM
MDC P/N Gemini EM Spectrofluorometer
SpectraMax M2
SpectraMax M5
FlexStation
®
®
MDC P/N M2
MDC P/N M5
MDC P/N FlexStationII Scanning Fluorometer
Embla 96/384 plate washer
MDC P/N 0200-3948
AquaMax DW4 with 96-well
MDC P/N AquaMax DW4 Liquid Handling System
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
4
CatchPoint cGMP Fluorescent Assay Kit Experimental
Protocol
Preparing the
Reagents
Table 4: Preparation of reagents
Step Action
1
Reconstitute one vial of 30 µM cGMP Calibrator:
Add 5 mL of cGMP Assay Buffer to the vial. Mix well
to ensure dissolution of all contents. Store on ice or at 4°C.
2 Reconstitute one vial of HRP-cGMP Conjugate:
Add 10 mL of cGMP Assay Buffer to one of the vials for a 10X stock.
Mix well to ensure dissolution of all contents. Transfer the entire
contents of the vial to a 100 mL graduated cylinder by repeated
washings with assay buffer. Adjust the final volume to 100 mL with
assay buffer and mix thoroughly. The diluted HRP-cGMP conjugate is
ready for use in the assay. Store on ice or at 4°C.
3 Reconstitute one vial of Rabbit anti-cGMP Antibody:
Add 10 mL of cGMP Assay Buffer to one of the vials for a 10X stock.
Mix well to ensure dissolution of all contents. Transfer the entire
contents of the vial to a 100 mL graduated cylinder by repeated
washings with assay buffer. Adjust the final volume to 100 mL with
assay buffer and mix thoroughly. The diluted Rabbit anti-cGMP
conjugate is ready for use in the assay. Store on ice or at 4°C.
4 Dilute 10X Wash Concentrate 10-fold in deionized
water, i.e. for one assay plate dilute 500 mL of 10X Wash Concentrate
in 4500 mL of deionized water.*
5
Prepare calibrators: The 30 µM cGMP calibrator is diluted in cGMP
Assay Buffer to prepare stock calibrators of 10000, 100, 33, 11, 3.7,
1.2, and 0.4 nM cGMP. These give 3300, 33, 11, 3.7, 1.2, 0.41, and
0.14 nM (400, 4.0, 1.3, 0.44, 0.15, 0.049, and 0.016 pmol) final
concentrations in the assay.
1. Add 500 µL of the 30 µM cGMP calibrator to 1000 µL of cGMP
Assay Buffer. This is the 10,000 nM stock calibrator.
2. Add 20 µL of the 10,000 nM stock calibrator to 1980 µL of cGMP
Assay Buffer. This is the 100 nM stock calibrator.
3. In cGMP Assay Buffer, serially dilute the 100 nM stock calibrator in
a 3-fold fashion five times to produce the 33, 11, 3.7, 1.2, and 0.4
nM cGMP stock calibrators. 40 µL of each calibrator are required
per replicate. Prepare 1000 µL of each (ample for 10 replicates).
As a 0 nM (zero dose) calibrator, use the cGMP Assay Buffer itself.
The following protocol provides sufficient reagents to run five 96-well plates, each with 16
calibrators, 2 controls, and up to 78 samples.
* On dilution of the 10X Wash Buffer concentrate, the reagent contains 0.02 M Tris, 150 mM
NaCl, 0.05% Tween 20, and 0.05% Proclin 200 (pH 7.4).
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
5
Preparing the
Plate
For ease of analysis and calculation, the assay plate contains the calibrators, controls,
and samples as outlined below. Use the suggested template (Table 5) and run the
samples in duplicate.
Note: The order of addition of each component to the wells and the incubation time are
important. Mix samples and all reagents gently but thoroughly before use.
All assay wells will be prepared by dispensing the assay components in the following order:
1) Calibrator, control, or sample
2) Rabbit anti-cGMP Antibody
3) HRP-cGMP Conjugate
Table 5: To prepare the assay plate:
Step Action
Designing your assay plate (plate map):
1
We suggest preparing the plate according to the template below. If you
are using SoftMax® Pro to analyze your results, you may set up a
template before or after reading the plate. If you set up the plate before,
you can print out a template to help in preparing your plate.
Column
Row1 2 3
A 3300 nM 3300 nM Control 1
(400 pmol) (400 pmol)
B 33 nM 33 nM Control 2
(4 pmol) (4 pmol)
C 11 nM 11 nM
(1.3 pmol) (1.3 pmol)
D 3.7 nM 3.7 nM
(0.44 pmol) (0.44 pmol)
E 1.2 nM 1.2 nM Column 3 (C –H) and
(0.15 nM) (0.15 nM) all of columns 4 - 12:
F 0.41 nM 0.41 nM Samples for analysis
(0.049 pmol) (0.049 pmol)
G 0.14 nM 0.14 nM
(0.016 pmol) (0.016 pmol)
H 0 nM 0 nM
(buffer only) (buffer only)
2
Add 40 µL of the appropriate concentration of Calibrator working
solution (see step 5 from “Preparing the Reagents” on Table 4, page 4)
to columns 1 and 2 (rows A – H). Column 3 (A & B) is designated here
for optional controls. For example, one may wish to use a zero dose
calibrator with no Rabbit anti-cGMP Antibody to evaluate background.
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
6
Table 5: To prepare the assay plate:
3
Place 40 µL of samples to be analyzed in appropriate wells [e.g.
Column 3 (C – H), and columns 4 – 12]. See Appendix A for cell
stimulation and lysis protocols.
4
Add 40 µL of reconstituted Rabbit anti-cGMP Antibody to all wells
except any reserved for no antibody controls. If these are to be
performed, add 40 µL assay buffer in place of antibody.
Place plate on shaker for 5 minutes or gently agitate by hand to ensure
5
mixing.
6
Add 40 µL reconstituted HRP-cGMP Conjugate to every well.
Mix well and allow to incubate 2 hours at room temperature.
7
8
Aspirate plate contents and wash 4 times with 300 µL wash buffer
for each wash.
9
Prepare Stoplight Red Substrate: Dilute 600 µL of 100X stock Stoplight
Red Substrate into 60 mL of Substrate Buffer, then add 68 µL of 3%
H
2
0
2
(880 mM) to bring the final concentration to 0.0034% (1 mM) H
2
0
2
.
Important: The reconstituted substrate solution is sensitive to light. To
insure optimal performance we recommend preparing a fresh stock
solution and adding directly to assay plate within 60 min. At all times
keep substrate protected from light.
10
Add 100 µL Stoplight Red substrate to every well, minimizing the time
between starting and finishing. Cover the plate and leave at room
temperature for at least 10 minutes, shielded from light.
Read the fluorescence intensity of the plate on an appropriate
11
instrument (see below for settings). Plates may generally be read
anytime between 10 minutes and 24 hours. For optimal performance
you may want to briefly mix the plate prior to reading.
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
7
Analyst Parameter Reading the Plate in Analyst HTS Assay Detection System
Settings
The following instrument settings are recommended when using 96-well plates with a
100 µL volume in Analyst HTS systems:
Table 6: Fluorescence intensity settings for Analyst HT or GT Detection System:
Parameter
Setting
Fluorescence Intensity
530-25 nm
590-20 nm
Mode
Excitation filters
Emission filters
Dichroic mirror 50 / 50 beamsplitter
Z-height 3 mm
Attenuator Medium
Integration time
50,000 µsec
Lamp Continuous
Readings per well
PMT setup
Units
One
Smart Read+ (sensitivity = 2)
Counts / sec
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
8
Reading the Plate in the Gemini XS, Gemini XPS, Gemini EM, SpectraMax
®
M2 and
SpectraMax
®
M5 Detection System
The following instrument settings are recommended when using 96-well plates with a
Gemini XS system:
Table 7: Fluorescence intensity settings for GEMINI XS, Gemini XPS, Gemini EM,
SpectraMax
®
M2 and SpectraMax
®
M5 Detection System
Parameter Setting
Read Mode
Endpoint
Fluorescence (RFUs)
530 nm
590 nm
570 nm
6
Auto
On
Read Type
Excitation
Emission
Cutoff
Sensitivity
PMT
AutoCalibrate
Data Analysis
Analyzing the Calibration Data
The average intensity values (y-axis) can be plotted against calibrator concentration (x-axis)
to create a calibration or standard curve. Unknown samples can be interpolated using this
curve to calculate cGMP doses in the samples. We recommend using SoftMax Pro software
for accurate curve fitting and interpolation of data from MDC instruments. Figure 2 shows a
SoftMax Pro template and calibration curve for the cGMP assay when read on the Analyst™.
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
9
Figure 2. Calibration curve for the CatchPoint cyclic GMP Fluorescent Assay Kit, read on the Analyst system.
Data was taken 60 minutes after addition of Stoplight Red substrate. Fluorescence intensity settings for Analyst
HTS Detection System were as stated in Table 6. The EC
50
value of the calibration curve, calculated as
coefficient C using SoftMax Pro, was 2.2 nM (264 fmol).
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
10
ADDITIONAL INFORMATION
The following typical information is based on data from an Analyst HT. Absolute readings may differ on other
instruments, but relative values should be comparable.
Sensitivity (Typical values)
Minimum Detectable Concentration = 0.2 nM (24 fmol / well)
Maximum Measurable Concentration = 120 nM (14 pmol / well)
EC
50
= 2.5 + 0.6 nM (300 fmol)
Intra-assay precision
Intra-assay precision was calculated from measurements of 0 and 1.2 nM calibrator. The results
for a single assay are shown below:
Signal intensity / background
(for 0 nM calibrator)
§
%B/B
o
* precision for 1.2 nM calibrator
Signal intensity
Avg
/ background = 290
Standard deviation (SD) = 12
%CV = 4
n = 23
%B/B
o
(mean) = 59
Standard deviation (SD) = 4
%CV = 7
n = 23
§
Background taken as the signal intensity for wells in which no cGMP-HRP or Rabbit anti-cGMP
Antibody are added
* %B/B
o
= 100 X (I
1.2 nM
– I
∞
) / (I
o
– I
∞
)
where I =
1.2 nM
I =
o
I
=
∞
Signal intensity for 1.2 nM calibrator
Signal intensity for 0 nM calibrator
Signal intensity for
3333 nM calibrator
Inter-assay precision (three assays performed on separate days)
Inter-assay precision was calculated from measurements of %B/B
o
for 1.2 nM calibrator in 3 successive assays.
The results are shown below:
Measurement
Data points
%B/B
0
58
61
59
SD
5
4
4
%CV
8
6
7
A
B
C
n = 2
n = 2
n = 23
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
11
Precision correlation
22 replicates of each of the standards were prepared and % coefficient of variation was determined for each
concentration using SoftMax Pro.
Sample Concentration Mean Value
SD %CV
cal 01 3300 nM
2084707826
3.8
cal 02 33 nM
29
6.8
cal 03 11 nM
4876559202898
4.2
cal 04 3.7 nM
18
4.4
cal 05 1.2 nM
17
4.7
cal 06 0.41 nM
22
4.5
cal 07 0.14 nM
243627261233519
5.1
cal 08 0 nM
285929641049992
3.7
Appendix A: Preparing cell lysates for the cGMP Assay
The protocol below illustrates an application of the CatchPoint cGMP Fluorescent Assay. The protocol has
been developed for use with adherent cells stimulated with atrial natriuretic peptide (ANP). The procedure may
be modified for various hormones and stimulators, such as acetylcholine, insulin, serotonin, and nitric oxide.
For best results each cell line should be evaluated to determine the optimal concentration of cells per well,
incubation times, and concentration of guanylate cyclase agonists or antagonists.
Use this protocol as a guide to help optimize the assay.
______________________________________________________________________________________
Reagents
Reagent Stock Concentration
cGMP stimulator atrial natriuretic peptide,
300 µM
dissolved in phosphate buffered saline (PBS)
Phosphodiesterase inhibitor, 3-isobutyl-1-800 mM
methylxanthine (IBMX) dissolved in DMSO
Buffers
Buffer Source
Krebs-Ringer Bicarbonate Buffer (KRBG) Sigma Chemical Company (P/N K4002)
containing 10 mM Glucose (pH 7.4)
Note: 15 mM sodium bicarbonate is not
included with this product and must be
added separately.
Lysis Buffer (pH 7.3) Molecular Devices Corporation (included in
CatchPoint cGMP Fluorescent Assay Kit)
Phosphate Buffered Saline, PBS
Invitrogen Life Technologies
(P/N 10010-023)
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
12
Assay Protocol
Adherent Cells
The following has been used for the adherent cell lines RFL-6 and CHO-K1 :
1) Culture cells (1 mL / well) in standard 12-well microtiter plates (tissue culture grade) with cell concentration
at 0.75 – 2.5 x 10
5
cells/mL.
2) Incubate plated cells overnight at 37°C in a humidified atmosphere of 5% CO
2
: 95% air.
3) Gently aspirate media and slowly add 400 µL of Pre-stimulation Buffer (pH 7.4).
Pre-stimulation Buffer (containing 0.75 mM IBMX in KRBG Buffer; make fresh on day of experiment)
10 ml of KRBG Buffer (pH 7.4)
9.4 µL of 800 mM IBMX
Note: Once IBMX has been added to the KRBG, mix vigorously to obtain a homogeneous solution.
4) Incubate for 10 min at room temperature.
5) Add 200 µL of 3X ANP or PBS (negative control). Gently mix and incubate at 37°C for 15 minutes.
ANP, 3X (500 nM final concentration)
2.0 mL PBS (pH 7.4) containing
10 µL of 300 µM ANP (1.5 µM stock)
6) Add 200 µL of lysis buffer to each well. Agitate cells to facilitate cell lysis. This can be achieved by shaking
the plate on a plate shaker for 10 min after adding the lysis reagent.
7) If desired, carry out microscopic evaluation using Trypan Blue to check if cells have lysed (1:1 ratio). Cell
membrane may still be visible after cell lysis. Lysed cells are now ready for use and should be immediately
processed in the cGMP Fluorescent Assay Kit. Use 40 µL per well neat or suitably diluted lysate in
additional lysis buffer (see page 6, Table 5, Step 3 of the Product Insert).
8) Proceed with the remainder of the procedure in Table 5.
Cells in Suspension
For non-adherent cells, we recommend centrifugation of the cells from culture medium and suspension of the
pellet in Pre-stimulation Buffer (pH 7.4). Add 1 mL (12 well plate) of cell suspension to each well of the plate.
It is recommended that you then centrifuge the plates at 1000 rpm for up to 4 min (brake off). Proceed with step
3 (shown above) for adherent cells.
R3368B
CatchPoint Cyclic-GMP Bulk Fluorescent Assay Kit_96-well format
13
版权声明:本文标题:CatchPoint Cyclic-GMP Fluorescent Assay Kit产品说明书 内容由网友自发贡献,该文观点仅代表作者本人, 转载请联系作者并注明出处:http://roclinux.cn/p/1735538742a1674296.html, 本站仅提供信息存储空间服务,不拥有所有权,不承担相关法律责任。如发现本站有涉嫌抄袭侵权/违法违规的内容,一经查实,本站将立刻删除。
发表评论